ValleyLab FX

The Valleylab Force FX™ is an isolated output electrosurgical generator that provides the power for cutting, desiccating, and fulgurating tissue during surgery. The generator is used for bipolar and monopolar electrosurgical procedures.

Categories: ,

Description

You can use handset or footswitch controls to activate the generator. Receptacles are marked to ensure the correct connection of accessories. Indicators alert you when the generator is activated.

The Force FX™ allows you to adjust the activation tone volume and reset the mode and power settings used previously. An RF activation port, RS-232 serial port, and expansion port are also included.

The Force FX™ can be used in conjunction with the Valleylab Force GSU System and the Valleylab Electroshield Monitor. Refer to the manuals provided with these products for more information.

OUTPUT WAVEFORMS

  • BIPOLAR
  • PRECISE: 470 KHZ sinusoid
  • STANDARD: 470 KHZ sinusoid
  • MACRO: 470 KHZ sinusoid
  • MONOPOLAR CUT
  • LOW: 390 KHZ sinusoid. Similar to the pure cut mode except the maximum voltage is limited to a lower value.
  • PURE: 390 KHZ sinusoid
  • BLEND: 390 KHZ bursts of sinusoid, recurring at 27 KHZ intervals
*50% duty cycle envelope

MONOPOLAR COAG

DESICCATE: 240 KHZ sinusoid repeated at 39 KHZ. 8% duty cycle.
FULGURATE: 390 KHZ damped sinusoidal bursts witha randomized repetition centered at 28 KHZ. Frequencies include 21 KHZ.
*Output power changes by less than 15% or 5 watts, whichever is greater, as the line voltage varies from 104-132 volts and 208 – 264 volts (at rated load).

LOW FREQUENCY LEAKAGE (50 – 60 HZ)

  • Source current, patient leads, all outputs tied together.
  • Normal polarity, intact chassis ground < 10 µA
  • Normal polarity, ground open< 50 µA
  • Reverse polarity, ground open< 50 µA
  • Sink current, 140V applied, all inputs < 50 µA

PER = 98

Per (power efficiency rating) is the measure of an electrosurgical generator’s ability to accurately deliver the selected power into a wide range of tissue types.

WEIGHT AND DIMENSIONS

  • HEIGHT: 11.1 CM (4 3/8″)
  • WIDTH: 35.6 CM (14″)
  • LENGTH: 43.9 CM (17″)
  • WEIGHT: < 8.1 KG (< 18 LBS)

INPUT POWER REQUIREMENTS:

  • Operating range is 85 to 132 ac volts. Maximum current is 7 amperes in cut and 4 amperes in coag.

HIGH FREQUENCY LEAKAGE

  • BIPOLAR: Less than 60 M ARMS
  • MONOPOLAR: Less than 150 M ARMS

REM™ CONTACT QUALITY MONITORING SYSTEM

  • MEASUREMENT FREQUENCY: 80 KHZ ± 10 KHZ
  • MEASUREMENT CURRENT: Less than 10 µA
  • ACCEPTABLE RESISTANCE RANGES:
  • REM™ PAD – 5-135 OHMS
  • NON-REM™ PAD – Less than 20 OHMS
  • Acceptance range is 5-135 OHMS after Rem Polyhesive™ II return electrode is applied.

ADAPTIVE REM
Rem™ T rip is baseline impedance plus 40%. For example, if the baseline impedance is 30 ohms, the upper level trip is approximately 42 OHMS. If the pad-patient impedance falls below the baseline impedance, a new baseline is established.

OUTPUT CHARACTERISTICS
MODE MAXIMUM P-P VOLTAGE RATED LOAD (OHMS) MAXIMUM POWER (WATTS) CREST FACTOR (TYPICAL)
BIPOLAR PRECISE 450 100 70 1.5
STANDARD 320 100 70 1.5
MACRO 750 100 70 1.5
MONOPOLAR CUT LOW 1350 300 300 1.5
PURE 2300 300 300 1.5
BLEND 3300 300 200 2.5
MONOPOLAR COAG DESICCATE 3500 500 120 5.0
FULG. HIGH CRESTFACTOR 8500 500 120 7.0
FULG. LOW CREST FACTOR 6900 500 120 5.5
SPRAY 9000 500 120 8.0
CEM MODE MONOPOLAR CUR (LOW) 1000 300 100 1.5
MONOPOLAR COAG (DESICCATE) 3500 500 70 5.0

*Crest Factor ia an indicator of a waveform’s ability to coagulate without cutting.

  • Click Here To Request A Quote