ValleyLab SSE2L

The Valleylab SSE2L is designed to provide the three most significant electrosurgical effetcs:

  • cutting
  • desiccation
  • fulguration

*Provides both Monopolar and Bipolar outputs for electrosurgery – Preset blended cutting modes provide optimum hemostasis

Categories: ,

Description

  • The SSE2L allows up to 350 watts output on a cut setting
  • A monopolar foot switch is included
  • Refurbished SSE2Ls are available from DRE

Optional:

  • REM™ (Return Electrode Monitoring)
  • Bipolar foot switch is optional
  • Rolling Cart  

Specifications

Dimensions: 7.6″H x 11.6″W x 16.6″D | 19.3cm x 29.5cm x 42.2cm

Weight:

  • 100/110 V – 32.5 lbs. (14.7 kg)
  • 200/240 V – 36.5 lbs. (16.6 kg)

Maximum Wattage: 350 watts

Output Waveform:

  • Cut: 500 kHz ± 10 kHz sinusoid with 65% 120 Hz modulation
  • Blend: 500 kHz ± 10 kHz sinusoid with increased amplitude bursts at 20 kHz
  • Coag: Pulse modulated 450 kHz damped sinusoid. Burst duration 7.0 µSec, nominal. Burst repitition 20 kHz ± 2 kHz
  • Bipolar: 500 kHz ± 10 kHz sinusoid, unmodulated

Monopolar Output Characteristics:

Mode Maximum (Open Circuit P-P Voltage Rate Load (Ohms) Control Setting = 10 (Rated Load)(Watts) Nominal Power Crest Factor @ Rated Load (typical)
Pure Cut 2800 500 375 ± 25 1.5 CS = 5
Blend 2440 500 250 ± 25 2.4 CS = 7
Coag 4070 500 125 ± 15 6.4 CS = 10
(3500 220/240V)

 

Bipolar Output Characteristics:

Mode Maximum (Open Circuit P-P Voltage Rate Load (Ohms) Control Setting = 10 (Rated Load)(Watts) Nominal Power Crest Factor @ Rated Load (typical)
Pure Cut 700 100 90+15-20 1.8 CS = 5
Blend 640 100 85+15-20 2.1 CS = 5
Coag 1100 100 25+5-10 5.1 CS = 5
(900 220/240V)

Input Power Source: (Control setting of 10, outpu short circuit)

90-135 VAC, 50/60 Hz 180-270 VAC, 50/60 Hz
Idle: 0.6 Amperes Idle: 0.3 Amperes
Cut: 12 Amperes Cut: 6 Amperes
Coag: 4 Amperes Coag: 2 Amperes

 

 

  • Click Here To Request A Quote